    Perbedaan Nilai Activated Partial Thromboplastin Time (Aptt) & Prothrombin Time (Pt) Metode Point-Of-Care Testing (Poct) Dan Electromechanic Semi Automatic

    Differences In Activated Partial Thromboplastin Time (Aptt) & Prothrombin Time (Pt) Methods Point-Of-Care Testing (Poct) And Electromechanic Semi Automatic

    Thumbnail
    View/Open
    Cover (666.0Kb)
    Fulltext (3.058Mb)
    Date
    2024
    Author
    Syifa, Mutiara Ayu
    Advisor(s)
    Siregar, Dewi Indah sari
    Arto, Nindia Sugih
    Metadata
    Show full item record
    Abstract
    BACKGROUND: Currently, there are several methods of coagulation testing, one of which is the Point-of-care testing (POCT) method which has emerged as a tool that plays an important role in modern healthcare, enabling fast, convenient diagnosis as well as more economical costs and can be done directly at the patient's treatment site. Some laboratories have also used the semi-automatic elecromechanic coagulation analyser as a coagulation testing tool. OBJECTIVES: To see the difference of aPTT and PT values in POCT method and semi automatic electromechanic method, so that it is useful in the selection of coagulometer. METHODS: This study is an observational study with a cross sectional design conducted at H. Adam Malik Hospital Medan. Using 100 samples, each sample was examined for aPTT, PT and INR using the POCT and electromechanic semi-automatic methods. The results of each examination were analysed using the Mann Whitney statistical test. RESULTS: The results of the comparative test between INR of the POCT method and electromechanic semi-automatic found a significant difference between the two examination methods (p = <0.001). While in the aPTT an PT parameter, the POCT method with semi-automatic electromechanic did not find a significant difference between the two examination methods (p = 0.434 and p= 0.371). The average value of the difference between the two tools was 4.12 seconds in aPTT parameters and 0.50 in INR. CONCLUSIONS: There is no significant difference in the results of the aPTT and PT parameter between the two methods but there is a difference in the results of the INR parameter.
    URI
    https://repositori.usu.ac.id/handle/123456789/108908
    Collections
    • Master Theses (Clinical Medicine) [1506]

    Repositori Institusi Universitas Sumatera Utara - 2025

    Universitas Sumatera Utara

    Perpustakaan

    Resource Guide

    Katalog Perpustakaan

    Journal Elektronik Berlangganan

    Buku Elektronik Berlangganan

    DSpace software copyright © 2002-2016  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    Atmire NV
     

     

    Browse

    All of USU-IRCommunities & CollectionsBy Issue DateTitlesAuthorsAdvisorsKeywordsTypesBy Submit DateThis CollectionBy Issue DateTitlesAuthorsAdvisorsKeywordsTypesBy Submit Date

    My Account

    LoginRegister

    Repositori Institusi Universitas Sumatera Utara - 2025

    Universitas Sumatera Utara

    Perpustakaan

    Resource Guide

    Katalog Perpustakaan

    Journal Elektronik Berlangganan

    Buku Elektronik Berlangganan

    DSpace software copyright © 2002-2016  DuraSpace
    Contact Us | Send Feedback
    Theme by 
    Atmire NV