Analisis Variasi Ekspose terhadap Kontras Jaringan Lunak (Soft Tissue) dengan Menggunakan Radiografi
View/ Open
Date
2014Author
Pangaribuan, Herlina Maryani
Advisor(s)
Warman, Aditia
Humaidi, Syahrul
Metadata
Show full item recordAbstract
It had been research to measure factor exposition to soft network base used X-ray
craft with variation of tube tension (kV) and Current (mAs). Experiment is
conducted with object irradiating soft tissue, at condition of tube tension
variation 42 kV, 46 kV, 52 kV, 57 kV and 63 kV with current fixed namely 8
mAs. Then assessed density by using densitometer. Highest contrast for
variation of tube tension are 42 kV as high as 0.69, Contrast lowest value for
variation of tube tension namely 63 kV as high as 0.06.
And then research continue with current variation 2 mAs, 4 mAs, 7 mAs, 8 mAs
and 10 mAs with tube tension fixed namely 46 kV. Whereas highest contrast
value for current variation is 4 mAs as high as 1.15, contrast lowest for
current variation is 8 mAs as high as 0.30.