Optimasi Citra Radiografi pada Pemeriksaan Pelvis Menggunakan Computed Radiography (CR)
View/ Open
Date
2013Author
Wahyudi, Ridho
Advisor(s)
Sembiring, Kurnia
Metadata
Show full item recordAbstract
A research about a radiography image optimization using a rando phantom
by Computed Radiography (CR) type Prima T Drypic 2000 system has been
done, for examination of Antero Posterior (AP) pelvis. The optimization of image
formation was evaluated based on guidance from European Commission with
their acceptance criterian, the condition of kVp and mAs, high contrast and low
contrast. Beside the image visual evaluation for optimization, the Computed
Radiography (CR) incident exposure was also observed which is stated in
Sensitivity Value (S Value) by image digitations process that can be seen at image
console and imageJ software. The result of optimization research show that for the
AP pelvis, optimization was when 75 kVp 10 mAs