Ansalisis Kualitas Citra Brain Variasiflipangle(FA) dengan Menggunakan Squens Spin Echo Pada Magnetic Resonance Imaging (MRI)
View/ Open
Date
2014Author
Rajagukguk, Rudi Hasudungan
Advisor(s)
Sembiring, Manis
Humaidi, Syahrul
Metadata
Show full item recordAbstract
This Research done at image three people of patient by using MRI 1,5T
inspection of MRI Brain, to analyse influence of FA to quality of image of MRI
brain of wight of T2 spin echo. items of Research result of image of MRI obtained
from variation of flip agle, wight of T2 spin echo to quality of image Brain, what
is done in each patient. the Research in RS Field Firmness. Procedures of
Research with preparation of Magnetic implement of Resonance Imaging, patient
preparation, arrangement from Flip Angel. Result of statistical test of this
research is obtained by that FA 1200
,1400
and 1600 having an effect on signifikan
to quality draw, with value of TR of 5000 of ms and TE of 96 ms of its meaning is
of FA to quality draw contrast of image of Brain cutting of axial with analysing
result of Middle measurement, Under right, Under Left, For Right and For Left
getting that middle area more signifikan because more amount is aqueous. Than
8 image assessed to own quality of more optimal image at flip angle ( FA) 1600
.
From result obtained inferential that progressively go up value of Flip Angle (
FA) hence will progressively improve Signal Noise to Ratio (SNR).