dc.description.abstractTesting and calibration of INCUBATOR BABY YP-930 has been performend using incu analyzer, safety analyzer 601 and held at Balai Pengamanan Fasilitas Kesehatan Republik Indonesia, with the proviso that the INCUBATOR output energy is in the threshold of ± 10ºC temperature, and ≤ 400ºC temperature Mats, and ≤ 65 dBA Noise, ± 0,35 m / second temperature speed of setting at the time of testing. From the result of research conducted has ben know that the INCUBATOR BABY YP- 930 wich has been tested is feasible to use.en_US