Pemanfaatan Cesium-137 pada Uji Homogenitas Dosis dan sebagai Sumber Standar Kalibrasi Alat Ukur Radiasi Dosimetri Radiation Detector Bosean Fs-600
dc.contributor.advisor | Maghfirah, Awan | |
dc.contributor.advisor | Rahim, Abdul | |
dc.contributor.author | Harefa, Rabiatul Adawiyah | |
dc.date.accessioned | 2025-01-03T03:29:20Z | |
dc.date.available | 2025-01-03T03:29:20Z | |
dc.date.issued | 2024 | |
dc.identifier.uri | https://repositori.usu.ac.id/handle/123456789/99764 | |
dc.description.abstract | A study on the utilization of Cesium-137 as a standard calibration source for the BOSEAN FS-600 Radiation Detector has been conducted. The research aimed to evaluate the stability of the dosimeter system, determine the calibration factor, and test dose homogeneity using the BOSEAN FS-600 Radiation Detector. The calibration process was carried out at varying distances of 100 cm, 125 cm, 150 cm, 175 cm, and 200 cm. Dosimeter stability was assessed using a control chart method, with a standard deviation of 0.115%. Calibration measurements were performed with a Cesium-137 radiation source at an ambient dose equivalent rate H(10) of 567.156912 μSv/Minute, resulting in an average calibration factor of 1,008±1,92. This value falls within the international tolerance range (0.8–1.2). Furthermore, the measured radiation distribution was 6.43%, indicating good dose homogeneity across the distance variations. These results suggest that Cesium-137 can be relied upon as a standard calibration source, providing accurate data in compliance with occupational safety standards. This study offers a significant contribution to the advancement of precise radiation measuring instruments, applicable in medical, industrial, and environmental protection contexts to mitigate uncontrolled radiation exposure. | en_US |
dc.language.iso | id | en_US |
dc.publisher | Universitas Sumatera Utara | en_US |
dc.subject | BOSEAN FS-600 | en_US |
dc.subject | Cesium-137 | en_US |
dc.subject | Calibration | en_US |
dc.subject | Dosimeter | en_US |
dc.subject | Dose Homogeneity | en_US |
dc.title | Pemanfaatan Cesium-137 pada Uji Homogenitas Dosis dan sebagai Sumber Standar Kalibrasi Alat Ukur Radiasi Dosimetri Radiation Detector Bosean Fs-600 | en_US |
dc.title.alternative | Utilization of Cesium-137 in Dose Homogeneity Testing and as a Standard Calibration Source for the Radiation Measuring Device Bosean Fs-600 Dosimeter | en_US |
dc.type | Thesis | en_US |
dc.identifier.nim | NIM200801032 | |
dc.identifier.nidn | NIDN0002097904 | |
dc.identifier.kodeprodi | KODEPRODI45201#Fisika | |
dc.description.pages | 89 Pages | en_US |
dc.description.type | Skripsi Sarjana | en_US |
dc.subject.sdgs | SDGs 9. Industry Innovation And Infrastructure | en_US |
Files in this item
This item appears in the following Collection(s)
-
Undergraduate Theses [1300]
Skripsi Sarjana